Irradiance and Dose Calculations Created by IPL Light in the Muscle Phantom

dc.contributor.authorDurmus, Huseyin Okan
dc.contributor.authorKaraboce, Baki
dc.contributor.authorSeyitsoy, MirHasan
dc.date.accessioned2025-10-29T11:13:57Z
dc.date.issued2021
dc.departmentFakülteler, Temel Bilimler Fakültesi, Fizik Bölümü
dc.descriptionMedical Technologies Congress (TIPTEKNO'21) -- NOV 04-06, 2021 -- Antalya, TURKEY
dc.description.abstractIn this study, the temperature effects caused by the IPL light source sent at regular intervals by a commercial epilation device on four thermocouples placed 15 mm depth inside the surface of the muscle phantom were measured at five different power levels formed for particular skin tones, and the safety evaluation of the IPL device in terms of temperature was made. In addition, the muscle phantom used in the study was characterized optically and acoustically. Furthermore, based on the emissivity data in the literature, the irradiance and light doses at a distance of 15 mm inside the muscle phantom were calculated. The calculated light doses were compared with the dose values applied for light therapies in the literature.
dc.description.sponsorshipBiyomedikal Klinik Muhendisligi Dernegi,Izmir Ekonomi Univ
dc.identifier.doi10.1109/TIPTEKNO53239.2021.9632893
dc.identifier.isbn978-1-6654-3663-2
dc.identifier.issn#DEĞER!
dc.identifier.orcid0000-0003-0973-0643
dc.identifier.scopus2-s2.0-85123727093
dc.identifier.scopusqualityN/A
dc.identifier.urihttps://doi.org/10.1109/TIPTEKNO53239.2021.9632893
dc.identifier.urihttps://hdl.handle.net/20.500.14854/7003
dc.identifier.wosWOS:000903766500011
dc.identifier.wosqualityN/A
dc.indekslendigikaynakWeb of Science
dc.indekslendigikaynakScopus
dc.language.isotr
dc.publisherIEEE
dc.relation.ispartofTip Teknolojileri Kongresi (Tiptekno'21)
dc.relation.publicationcategoryKonferans Öğesi - Uluslararası - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/closedAccess
dc.snmzKA_WOS_20251020
dc.subjectMuscle Phantom
dc.subjectIPL (Intense Pulsed Light)
dc.subjectEpilation Device
dc.subjectTemperature Impacts
dc.subjectIrradiance
dc.subjectDose Calculation
dc.titleIrradiance and Dose Calculations Created by IPL Light in the Muscle Phantom
dc.typeConference Object

Dosyalar