Energy resolution methods efficiency depending on beam source position of potassium clusters in time-of-flight mass spectrometer
Yükleniyor...
Tarih
Yazarlar
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
Indian Acad Sciences
Erişim Hakkı
info:eu-repo/semantics/closedAccess
Özet
Energy resolution of the time-of-flight mass spectrometer was considered. The estimations indicate that the time-lag energy focusing method provides better resolution for the parallel case while the turnaround time is more convenient for the perpendicular position. Hence the applicability of the methods used for the energy resolution depends on beam source arrangement.
Açıklama
Anahtar Kelimeler
energy resolution, time-lag energy focusing method, turnaround time method, time-of-flight mass spectrometer
Kaynak
Pramana-Journal of Physics
WoS Q Değeri
Scopus Q Değeri
Cilt
69
Sayı
3