Design of novel anthracene-based fluorescence sensor for sensitive and selective determination of iron in real samples

Yükleniyor...
Küçük Resim

Tarih

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Elsevier Science Sa

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

A dipodal fluorescence (FL) system based on an anthracene platform was prepared by a general synthetic strategy. The water-miscible receptor was characterized by MALDI-MS, H-1 NMR, C-13 NMR, and FT-IR. In order to the illustration of photophysical properties of the synthesized sensor, the excitation-emission matrix (EEM) analysis and 3D-fluorescence measurements were performed. Ultraviolet-Visible (UV-vis) and FL spectroscopies were carried out to evaluate the FL performance of the sensor. Low concentrations of iron ions (Fe3+) could be determined by this sensitive and simple sensor. Due to the quenching effect of Fe3+, a limit of detection (LOD) of 0.314 mu M was obtained in the linear range of 0.3-130 mu M. This fluorescent sensor was applied for measuring ferric ions in environmental water and tablet samples, which was highly efficient. The proposed strategy could be applied to develop sensors for not only for Fe3+ but also for many various metal ions using a diversity of fluorophores and ionophores.

Açıklama

Anahtar Kelimeler

Dipodal receptor, Anthracene, Fluorescent sensor, Environmental water, Tablet