Performance Optimization of a Microfluidic Virus Detection Cartridge: A Numerical and Experimental Study

dc.contributor.authorSenel, Enes Burak
dc.contributor.authorKizilelma, Bilal
dc.contributor.authorTamdogan, Enes
dc.contributor.authorYorulmaz, Mustafa
dc.date.accessioned2025-10-29T11:13:32Z
dc.date.issued2023
dc.departmentFakülteler, Temel Bilimler Fakültesi, Moleküler Biyoloji ve Genetik Bölümü
dc.description.abstractDetection and imaging of viruses in a complex solution is particularly significant for virology and requires a comprehensive understanding of biosensors. While lab-on-a-chip systems are used in virus detection as biosensors, analysis and optimization of these systems are especially challenging due to the size of the system to be used in the certain application. The system of interest for virus detection is required to be cost efficient and is also needed to be able to easily operable with a simple setup. Moreover, the detailed analysis of these microfluidic systems should be made with precision in order to predict the capabilities and the efficiency of the system accurately. This paper reports on the use of a common commercial computational fluid dynamics (cfd) software for the analysis of a microfluidic lab-on-a-chip virus detection cartridge. This study evaluates the problems commonly encountered during microfluidic applications of cfd softwares particularly in the area of reaction modeling of the antigen-antibody interaction. cfd analysis is later validated and combined with experiments to optimize the amount of dilute solution used in the tests. Thereafter, the geometry of the microchannel is also optimized and optimal test conditions are set for a cost efficient and effective virus detection kit using light microscopy.
dc.description.sponsorshipThe datasets generated and supporting the findings of this article are obtainable from the corresponding author upon reasonable request.
dc.description.sponsorshipThe datasets generated and supporting the findings of this article are obtainable from the corresponding author upon reasonable request.
dc.identifier.doi10.1115/1.4062850
dc.identifier.issn0148-0731
dc.identifier.issn1528-8951
dc.identifier.issue10
dc.identifier.orcid0000-0002-2015-9890
dc.identifier.pmid37382621
dc.identifier.scopus2-s2.0-85164845755
dc.identifier.scopusqualityQ3
dc.identifier.urihttps://doi.org/10.1115/1.4062850
dc.identifier.urihttps://hdl.handle.net/20.500.14854/6809
dc.identifier.volume145
dc.identifier.wosWOS:001059030900007
dc.identifier.wosqualityQ3
dc.indekslendigikaynakWeb of Science
dc.indekslendigikaynakScopus
dc.indekslendigikaynakPubMed
dc.language.isoen
dc.publisherAsme
dc.relation.ispartofJournal of Biomechanical Engineering-Transactions of the Asme
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/closedAccess
dc.snmzKA_WOS_20251020
dc.subjectbiosensors
dc.subjectmicrofluidics
dc.subjectvirus detection
dc.subjectcomputational fluid dynamics (CFD)
dc.subjectantigen
dc.subjectlight microscopy
dc.titlePerformance Optimization of a Microfluidic Virus Detection Cartridge: A Numerical and Experimental Study
dc.typeArticle

Dosyalar