A Novel Poly(Glutaraldehyde-co-Pyrrole)/Horseradish Peroxidase Composite Film Electrode

Yükleniyor...
Küçük Resim

Tarih

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Taylor & Francis Inc

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

A newly synthesized copolymer, Poly(glutaraldehyde-co-pyrrole) was used to immobilize horseradish peroxidase for the construction of an amperometric biosensor for detection of phenols in a flow injection system. Infrared spectra of Poly(glutaraldehyde-co-pyrrole) showed that the incorporation of aldehyde groups into the conductive polymeric backbone was achieved. The [image omitted] values were found to be in the range of 0.077-0.468mM for p-benzoquinone and 0.112-0.709mM for catechol, at various flow rates ranging between 0.25-6mL/min. The sensitivities at various flow rates ranged between 100-200nA/mM for catechol and 30-100nA/mM for p-benzoquinone. The biosensor retained its initial activity for 1 month.

Açıklama

Anahtar Kelimeler

Amperometric biosensor, p-benzoquinone, catechol, flow injection, poly(glutaraldehyde), poly(pyrrole)