Analog filters for Enhanced Signal Reception of Magnetic Particle Imaging (MPI) Scanner
Yükleniyor...
Tarih
Yazarlar
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
Institute of Electrical and Electronics Engineers Inc.
Erişim Hakkı
info:eu-repo/semantics/closedAccess
Özet
This paper presents the analog filters for the receive chain of the magnetic particle imaging (MPI) scanner. Direct feedthrough interference between drive and receive coil was evaluated with three analog filters. LTspice® simulation tool was used to design and simulate analog filters at 10 kHz. Maximum attenuation of-80 dB was achieved with the 7th order Chebyshev filter. As a result of the higher-order filter, the direct feedthrough effect was also minimized. The signal-To-noise ratio (SNR) would also be improved with a good receive section of the MPI scanner. © 2022 Elsevier B.V., All rights reserved.
Açıklama
2021 Medical Technologies Congress, TIPTEKNO 2021 -- Antalya -- 175481
Anahtar Kelimeler
field of view, field-free point., magnetic particle imaging